| Model No. |
TD-M500 |
| Measuring Technique |
Attenuated Light Method |
| Measuring Range |
0.0-500FTU |
| Display Resolution |
0.1FTU (0.0-99.9), 1FTU (100-500) |
| Calibration |
Distilled Water or Air Calibration |
| Correction |
Single or Double Point Correction |
| Power Requirements |
9V Layer-built Alkaline Dry
Battery (6LR61) |
| Battery Life |
Approx. 8000 measurements |
| Water Temperature |
0 to 40 (32 to 104F) |
| Operating Temperature |
-5 to 50 (23 to 122F) |
| Operating Humidity |
95% Rh or less |
| Materials |
Controller: Body / ABS, Optical
Windows / PMMA
Probe: Body / SUS316L, Protection Cover / Silicon Rubber |
| Cable Length |
5m |
| Dimensions |
Controller: H x W x D = 160
x 44 x 42 mm
Probe: 30 x116 mm |
| Weight |
0.6kg (Incl. Battery) |
| Protective Structure |
Controller: IP67
Probe: IP68 |